PMID- 31283813 OWN - Quintessenz Verlags-GmbH CI - Copyright Quintessenz Verlags-GmbH OCI - Copyright Quintessenz Verlags-GmbH TA - Int J Prosthodont JT - The International Journal of Prosthodontics IS - 1942-4426 (Electronic) IS - 0893-2174 (Print) IP - 4 VI - 32 PST - ppublish DP - 2019 PG - 352-354 LA - en TI - Comparison of the Trueness of Lithium Disilicate Crowns Fabricated From All-in-One and Combination CAD/CAM Systems LID - 10.11607/ijp.6031 [doi] FAU - Lee, Jin-Ju AU - Lee J FAU - Son, Keunbada AU - Son K FAU - Bae, Eun-Bin AU - Bae E FAU - Choi, Jae-Won AU - Choi J FAU - Lee, Kyu-Bok AU - Lee K FAU - Huh, Jung-Bo AU - Huh J CN - AB - Purpose: To examine and compare the trueness between a computer-aided design (CAD) file and the final milled prosthesis. Materials and Methods: For each included patient (n = 30), two lithium disilicate crowns were produced (n = 60) using two different CAD/computer-assisted manufacture (CAD/CAM) systems (allin- one type and combination type). To compare the trueness, the scanned internal surfaces of the fabricated prostheses were compared to the design files in CAD using 3D analysis software (Geomagic Control X, 3D Systems). Results: The mean ± standard deviation root mean square value for posterior teeth was 32.2 ± 9.4 in the combination type group and 43.8 ± 11.7 in the all-in-one type group, indicating higher trueness for the combination type (P < .05). Conclusion: This study has shown that the final milled prosthesis in combination type CAD/CAM systems had better trueness than the all-in-one system. AID - 850136